Description
FLIR A615 w/45° Lens Thermo Vision A Series 640×480 GigE -20+2000C FLIR Systems A615 Series 45°
Imaging & Optical Data | |
Lens identification | Automatic |
Thermal sensitivity/NETD | < 0.05°C @ +30°C (86°F) / 50 mK |
Focus | Automatic or manual (built in motor) |
Measurement | |
Accuracy | ±2°C or ±2% of reading |
Measurement analysis | |
Atmospheric transmission correction | Automatic, based on inputs for distance, atmospheric temperature and relative humidity |
Optics transmission correction | Automatic, based on signals from internal sensors |
Emissivity correction | Variable from 0.01 to 1.0 |
Reflected apparent temperature correction | Automatic, based on input of reflected temperature |
External optics/windows correction | Automatic, based on input of optics/window transmission and temperature |
Measurement corrections | Global object parameters |
Ethernet | |
Ethernet | Control and image |
Ethernet, standard | IEEE 802.3 |
Ethernet, connector type | RJ-45 |
Ethernet, type | Gigabit Ethernet |
Ethernet, communication | TCP/IP socket-based FLIR proprietary and GenICam protocol |
Ethernet, protocols | TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP, IGMP, ftp, SMTP, SMB (CIFS), DHCP, MDNS (Bonjour), uPnP |
Digital input/output | |
Digital input | 2 opto-isolated, 10–30 VDC |
Digital output, purpose | Output to ext. device (programmatically set) |
Digital output | 2 opto-isolated, 10–30 VDC, max 100 mA |
Digital I/O, isolation voltage | 500 VRMS |
Digital I/O, supply voltage | 12/24 VDC, max 200 mA |
Digital I/O, connector type | 6-pole jackable screw terminal |
Digital input, purpose | Image tag (start, stop, general), Image flow ctrl. (Stream on/off), Input ext. device (programmatically read) |
Power system | |
External power operation | 12/24 VDC, 24 W absolute max |
External power, connector type | 2-pole jackable screw terminal |
Voltage | Allowed range 10–30 VDC |
Environmental data | |
Storage temperature range | –40°C to +70°C (-40 to 158°F) |
Humidity (operating and storage) | IEC 60068-2-30/24 h 95% relative humidity +25°C to +40°C (77 to 104°F) |
EMC | • EN 61000-6-2:2001 (Immunity) • EN 61000-6-3:2001 (Emission) • FCC 47 CFR Part 15 Class B (Emission) |
Vibration | 2 g (IEC 60068-2-6) FLIR Systems A615 |
Physical data | |
Housing material | Aluminium |
Scope of delivery | |
Hard transport case or cardboard box, Thermal imaging camera with lens, Utility CD-ROM, Calibration certificate, Ethernet™ cable, USB cable (FLIR A615), Mains cable, Power cable (pig-tailed), Power supply, Printed Getting Started Guide, Printed Important Information Guide, User documentation CD-ROM, Warranty extension card or Registration card, 6-pole screw terminal (mounted on camera) |
Reviews